Human Leptin ELISA Kit (ab179884)
Key features and details
- One-wash 90 minute protocol
- Sensitivity: 4.65 pg/ml
- Range: 15.63 pg/ml - 1000 pg/ml
- Sample type: Cell culture supernatant, Cit plasma, EDTA Plasma, Hep Plasma, Plasma, Serum
- Detection method: Colorimetric
- Assay type: Sandwich (quantitative)
- Reacts with: Human
Overview
-
Product name
Human Leptin ELISA Kit
See all Leptin kits -
Detection method
Colorimetric -
Precision
Intra-assay Sample n Mean SD CV% Serum 8 5.1% Inter-assay Sample n Mean SD CV% Serum 3 3.5% -
Sample type
Cell culture supernatant, Serum, Plasma, Hep Plasma, EDTA Plasma, Cit plasma -
Assay type
Sandwich (quantitative) -
Sensitivity
4.65 pg/ml -
Range
15.63 pg/ml - 1000 pg/ml -
Recovery
Sample specific recovery Sample type Average % Range Serum 91 90% - 93% Cell culture media 102 101% - 104% Hep Plasma 98 88% - 105% EDTA Plasma 82 80% - 87% Cit plasma 113 107% - 123% -
Assay time
1h 30m -
Assay duration
One step assay -
Species reactivity
Reacts with: Human -
Product overview
As of January 16, 2020, Human Leptin ELISA kit (ab179884) has been re-developed. We have identified new recombinant monoclonal antibodies to provide improved performance and consistency.
Human Leptin ELISA kit (ab179884) is a single-wash 90 min sandwich ELISA designed for the quantitative measurement of Leptin protein in human serum, plasma and cell culture supernatants. It uses our proprietary SimpleStep ELISA® technology. Quantitate human Leptin with 4.65 pg/mL sensitivity.
SimpleStep ELISA® technology employs capture antibodies conjugated to an affinity tag that is recognized by the monoclonal antibody used to coat our SimpleStep ELISA® plates. This approach to sandwich ELISA allows the formation of the antibody-analyte sandwich complex in a single step, significantly reducing assay time. See the SimpleStep ELISA® protocol summary in the image section for further details. Our SimpleStep ELISA® technology provides several benefits:
-Single-wash protocol reduces assay time to 90 minutes or less
-High sensitivity, specificity and reproducibility from superior antibodies
-Fully validated in biological samples
-96-wells plate breakable into 12 x 8 wells strips
A 384-well SimpleStep ELISA® microplate (ab203359) is available to use as an alternative to the 96-well microplate provided with SimpeStep ELISA® kits. -
Notes
Leptin is a secreted protein factor produced by adipocytes. It regulates energy balance and body fat deposits. Leptin deficiency in humans and mice can cause in obesity.Acute low levels of leptin triggers increased appetite and chronic leptin deficiency can block the onset of puberty.
-
Platform
Microplate
Properties
-
Storage instructions
Store at +4°C. Please refer to protocols. -
Components 1 x 96 tests 10X Human Leptin Capture Antibody 1 x 600µl 10X Human Leptin Detector Antibody 1 x 600µl 10X Wash Buffer PT (ab206977) 1 x 20ml Antibody Diluent 4BI 1 x 6ml Human Leptin Lyophilized Recombinant Protein 2 vials Plate Seals 1 unit Sample Diluent NS (ab193972) 1 x 50ml SimpleStep Pre-Coated 96-Well Microplate (ab206978) 1 unit Stop Solution 1 x 12ml TMB Development Solution 1 x 12ml -
Research areas
-
Function
May function as part of a signaling pathway that acts to regulate the size of the body fat depot. An increase in the level of LEP may act directly or indirectly on the CNS to inhibit food intake and/or regulate energy expenditure as part of a homeostatic mechanism to maintain constancy of the adipose mass. -
Involvement in disease
Defects in LEP may be a cause of obesity (OBESITY) [MIM:601665]. It is a condition characterized by an increase of body weight beyond the limitation of skeletal and physical requirements, as the result of excessive accumulation of body fat. -
Sequence similarities
Belongs to the leptin family. -
Cellular localization
Secreted. - Information by UniProt
-
Alternative names
- FLJ94114
- LEP
- LEP_HUMAN
see all -
Database links
- Entrez Gene: 3952 Human
- Omim: 164160 Human
- SwissProt: P41159 Human
- Unigene: 194236 Human
Images
-
Standard Curve comparison between original human Leptin SimpleStep ELISA kit and redeveloped Simplestep ELISA kit. SimpleStep ELISA kit shows increased performance.
-
SimpleStep ELISA technology allows the formation of the antibody-antigen complex in one single step, reducing assay time to 90 minutes. Add samples or standards and antibody mix to wells all at once, incubate, wash, and add your final substrate. See protocol for a detailed step-by-step guide.
-
The Leptin standard curve was prepared as described in Section 10. Raw data values are shown in the table. Background-subtracted data values (mean +/- SD) are graphed.
-
The concentrations of Leptin were measured in duplicates, interpolated from the Leptin standard curves and corrected for sample dilution. Undiluted samples are as follows: serum 2.5%, plasma (citrate) 10%, plasma (edta) 10%, and plasma (heparin) 10%. The interpolated dilution factor corrected values are plotted (mean +/- SD, n=2). The mean Leptin concentration was determined to be 19,448 pg/mL in serum, 5,865 pg/mL in plasma (citrate), 9,377 pg/mL in plasma (edta), and 9,463 pg/mL in plasma (heparin).
-
The concentrations of Leptin were measured in duplicates, interpolated from the Leptin standard curves and corrected for sample dilution. Undiluted samples are as follows: cell culture media 50%. The interpolated dilution factor corrected values are plotted (mean +/- SD, n=2).
-
Interpolated dilution factor corrected values are plotted (mean +/- SD, n=2). The mean Leptin concentration was determined to be 7,463 pg/mL with a range of 836 – 15,878 pg/mL.